4.
Trouble-Shooting
Error Messages
Display Message
Explanation
Action
Searching for XY-reference position
The machine searches for the initial
reference positions
Wait for the search process to finish
Reference pos. not found, close
cover
Accutom-50 was started with the
cover open and did not search for
the reference position
Close the cover, wait for the
reference search to finish and press
START
Cover open!
again
You are trying to start the cutting/
grinding process or a dressing while
the cover is open
Close the cover and press START
Process stopped by open cover
again
You have opened the cover during a
cutting/grinding process or a
dressing process
Close the cover and restart the
interrupted process
Process running
You are trying to start two different
processes at the same time
Wait for the previous process to
finish and try again
Y-stop position outside cutting range
Y-stop position outside grinding
range
The set stop position is outside the
maximum range in the Y-direction.
(max. 105.0 mm)
Adjust Y-stop position.
Y-stop position less than Y-relative
You try to start a cutting process
when the relative Y-position is larger
than or equal to the Y-stop
Position the specimen holder
correctly and/or change the Y-stop
position.
Feed less than 90% of pre-set value The force has exceeded the force
limit (LOW, MEDIUM, HIGH) and the
feed speed is automatically reduced
Decrease the feed speed or increase
the force limit next time you cut the
same material
Cutting stopped, feed too low
The feed speed has been reduced to
less than 10% due to the force
exceeding the force limit
Decrease the feed speed, increase
the force limit or change the cut-off
wheel and re-start the cutting
process
X-relative must be negative before
start
In connection with the grinding to a
STOP POSITION the relative X-
position is positive
Position your sample correctly by
moving the sample holder to the right
X-relative must be less than stop
pos.
In connection with the grinding to a
STOP POSITION the relative X-
position is larger than the pre-set
stop position
Position your sample correctly by
moving the sample holder to the right
Too many cuts, only xx cuts possible The no. of cuts and/or the slice
thickness exceed the no. of possible
cuts
Re-position your sample or press
START
Removal too large, move X-table
again to allow for
automatic adjustment of the no. of
slices
In connection with the grinding mode
REMOVAL, the removal is set higher
than the possible movement of the
table
Reduce the amount of material to be
removed or clamp the sample
differently
Содержание Accutom-50
Страница 1: ...Manual No 15017001 Date of Release 15 11 2012 Accutom 50 Instruction Manual...
Страница 66: ...Accutom 50 Gebrauchsanweisung Handbuch Nr 15017001 Auslieferungsdatum 15 11 2012...
Страница 133: ...Accutom 50 Mode d emploi Mode d emploi no 15017001 Date de parution 15 11 2012...
Страница 199: ...Manual No 15017001 Date of Release 15 11 2012 Accutom 50 Spare Parts and Diagrams...
Страница 204: ......
Страница 205: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......