Accutom-50
Instruction Manual
6
Pull out the recirculation tank.
Fill the tank with 3.88 l of water and 120 ml Struers Additive. The
water level should be 5 mm below the edge of the front hole in
the tank cover.
Check that the cover is fitted securely into place in the
recirculation tank and push the drawer with the recirculation tank
into place again.
Check that the end of the inlet tube has dropped into position.
Recirculation Unit
IMPORTANT
Always ensure that there is sufficient water in the tank as the recirculation
pump will be damaged if it is running dry.
Note:
Changing of cooling water should be done at least once a month.
IMPORTANT
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive each time you refill with water.
Do not use any oil, petrol, or turpentine-based additives, only Struers
additive.