Accutom-50
Instruction Manual
32
3.
Maintenance
Clean the cutting chamber with a damp cloth. Do not use tap
water as you risk overflow in the recirculation cooling water tank.
Remember to remove all dirt particles from the grate.
Clean the specimen holder head and the clamps for the dovetail
feed.
Clean the flanges.
Clean the transparent cover with a damp cloth.
The cooling unit should be checked for cooling water after 8 hours
use or at least every week. The unit must be refilled if the flushing
pump cannot reach the cooling water. Remember to add Struers
Additive for Cooling Fluid: One part of Additive for 33 parts of water.
To check the concentration of additive, use a refractometer.
Concentration = Brix value. The concentration of additive should
always lie between 2.7 and 3.3 %. Add Struers Additive for Cooling
Fluid if the concentration is too low.
Daily Service
WARNING!
Do not use alcohol, acetone or similar solvents.
Checking the Recirculation Unit
IMPORTANT
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive each time you refill with water.
Содержание Accutom-50
Страница 1: ...Manual No 15017001 Date of Release 15 11 2012 Accutom 50 Instruction Manual...
Страница 66: ...Accutom 50 Gebrauchsanweisung Handbuch Nr 15017001 Auslieferungsdatum 15 11 2012...
Страница 133: ...Accutom 50 Mode d emploi Mode d emploi no 15017001 Date de parution 15 11 2012...
Страница 199: ...Manual No 15017001 Date of Release 15 11 2012 Accutom 50 Spare Parts and Diagrams...
Страница 204: ......
Страница 205: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......