19 620 02 0117
62
Cleaning, Maintenance, Overhaul
5.1.4
Fuse replacement
The mains fuse holder is located on the rear of the D-ACTOR
®
200 between mains
connector and mains switch.
• Press the two tabs from the open spaces (Fig. 5-1/1 and 2) of the fuse holder
inwards and pull out the fuse holder of the housing.
1
2
1 u 2 open spaces to pull out the fuse holder
Fig. 5-1 Mains connector, fuse holder, mains switch
Fig. 5-2 Fuse holder
• Pull the old fuses out of the mains fuse holder.
• Replace the fuses (T 8 AL / 250 VAC).
• Push the mains fuse holder back into the opening until it engages.
5.2
Maintenance and safety checks
Preventive maintenance is not necessarily required. However, regular maintenance
may help to identify possible defects at an early stage and thus increase the safety and
service life of the device.
Maintenance services can be ordered from our regional representatives in your area or
directly from STORZ MEDICAL AG.
Independently of the national accident prevention regulations and test and inspection
intervals prescribed for medical devices, we recommend that functional checks (see
C
hapter
) and safety checks in accordance with MPBetreibV
(Germany), MPBV (Austria), MepV (Switzerland) be performed at least once a year.