System Description
3.5.3
Connecting the potential equalisation
At the rear side of the device a connection of a potential equalisation (Fig. 3-7/2) is
available if due to national regulations or room class a connection is required.
• Connect the cable for the potential equalisation to the PE connector of the
D-ACTOR 200 (Fig. 3-7/2) and connect it to your PE connector.
3.5.4
Connecting handpiece
• Insert the plug of the handpiece into the corresponding handpiece connector on
the left rear side of the device.
1 Handpiece connector
V-ACTOR
2 and 3 Handpiece connector
D-ACTOR
4 Connector VACU-ACTOR
3
1
2
4
Fig. 3-8 Connectors of the Handpieces and VACU-ACTOR
• Make sure that the red dot on the socket is aligned with the red dot on the
handpiece connector.
• Place the handpiece into the handpiece holder.
NOTE
Please also refer to the separate operating manual for your handpiece.
3.5.5
Connecting Tablet (optional)
1
1 USB connector to D-ACTOR 200
2 Main switch of the Tablet
2
Fig. 3-9 Tablet with charching device and USB cable