System Description
The mounting of the handpiece holder is equal for both types.
• Push the holder into the provided openings at the D-ACTOR
®
200.
3.5.1.2
Suction cup holder
1
1 Openings for insertion of the
handpiece and suction cup
holder
Fig. 3-6 Mounting the suction cup holder
• Push the holder into the provided openings at the D-ACTOR
®
200.
3.5.2
Connecting the electrical power supply
• Connect the supplied mains cable to the mains connector on the rear side of the
device.
1
1 Mains connector
2 Connector for
potential equalisation
2
Fig. 3-7 Connecting the electrical power supply
• Insert the mains cable into the socket.
ATTENTION !
When setting up the instrument, make sure that the air outlets on the housing of
the D-ACTOR
®
200 are not blocked.
The instrument must only be connected to properly earthed and correctly installed
shockproof sockets!
The device must be positioned in a way so that disconnection from the mains is
easy to do.