19 620 02 0117
63
Cleaning, Maintenance, Overhaul
The following checks should be performed to ensure that the D-ACTOR
®
200 operates
safely.
1 Earth leakage current test according to national regulations.
2 Earth impedance test (with mains cable, incl. applicator housing) according to
national regulations.
NOTE
For further details on content and performance of the safety checks please contact
your local dealer.
5.3
Disposal
When disposing of this medical product, no special measures have to
be observed. Please proceed in accordance with applicable country-
specific regulations. After expiration of its service life, dispose of the
D-ACTOR
®
200 as waste electronic equipment.
5.4
Repair
Repair work on defective devices must only be carried out by personnel suitably
authorised by STORZ MEDICAL. Only original STORZ MEDICAL spare parts may be
used for this purpose. The personnel suitably authorised can be from STORZ MEDICAL
or be representatives of STORZ MEDICAL agencies and dealers.
5.5
Service life
The average expected service life (MTTF) in accordance with EN60601-1:2005 +
A1:2012 is
– approximately 15 000 hours for the medical - electrical device D-ACTOR
®
200
– 100 cleaning and disinfection cycles for the suction cup
Please contact your local Storz Medical distributor in case of service.
For information about the service life of the other handpieces, please refer to the
separate operating manuals for the respective handpiece.
Exceeding the service life can be expected to result in a failure of the device and
accessories. This also applies to handpieces.
No warranty claims shall be accepted beyond the information given in
C
hapter
of the
D-ACTOR 200.