Instruction manual for Evario hospital bed
161
I. Visual Inspection
Ok
Not
Ok
Description of
Defect
What to Check...
Check for...
Visual Inspection of the Mechanical Components (if installed)
Type plate + warning labels on
bed
Available on bed frame,
legible
Patient lifting pole, location
sleeves
Grab handle with strap
No damage, deformation
Manufacturer's recommendation: Replace grab handle after 5
years
Bed frame: Mattress base,
chassis
No damage, deformation,
no split welded seams
Castors
No damage
Mattress base
No damage or deformation
Safety sides
No damage, cracks or de-
formation
Connecting elements
(screws, bolts, nuts, safety
caps)
Fixed position, complete-
ness
Wearing parts,
Movable joints
Plastic parts, various
No damage or severe wear
Free of sharp-edged dam-
age/breakage
II. Electrical Measurement according to DIN EN 62353 (VDE 07511):
Limit value
Measured
value
Resistance of earthing con-
ductor
Measuring point: PE pin on
the bed, head end.
Measuring
current
≥2 A
0.3 Ω
Ω
Leakage current of device,
direct/difference (place bed
with conductive castors in a
way that it is insulated).
0.5 mA
mA
1 Plug the bed mains cable
in the test socket on the
measuring instrument.
2 Connect the measuring
instrument probe to the
PE connector
(mattress base,
head end).
3 For the duration of the
measurements, activate
the motors using the
handset.
Содержание Evario
Страница 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Страница 125: ...Instruction manual for Evario hospital bed 125 22 22...
Страница 166: ...Instruction manual for Evario hospital bed 166 Notes...
Страница 167: ...Instruction manual for Evario hospital bed 167...
Страница 168: ...Australian Distributor...