Instruction manual for Evario hospital bed
109
11.3.2.1 Deleting the Maximum Mattress Base Height Setting
1 Repeat steps 1 - 2 from the Chapter
11.3.2. The OK ! shown on the display
confirms the change to the technician
control level.
2 The maximum mattress base height symbol
will appear on the display and the status
message X will be displayed for the deleted
function.
3 Press the DOWN button once. After press-
ing the button, the OK ! displayed indicates
the successful deletion of the mattress base
height restriction.
11.4 Restricting Minimum Mattress Base Height
A pre-set minimum mattress base height cannot be exceeded in the daily care routine or by the
patient. When the bed is lowered, it only moves to this position and then stops. The minimum mat-
tress base height must only be defined by technical staff or qualified medical staff who have been
trained by the operator to do so.
The mattress base must be horizontal before a restriction can be defined. An addi-
tional “LIMIT” notice on the display indicates an active limit during adjustment.
11.4.1 On the LCD Handset
1 Set the bed to the position that you wish to define as the minimum mattress base height.
2 Press the UP and DOWN buttons simul-
taneously, and at the same time hold the
top end of the handset briefly against the
orange magnetic unlocking key.
3 Keep both the UP and DOWN buttons
pressed for a further 5 seconds while the
warning triangle symbol is displayed.
Содержание Evario
Страница 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Страница 125: ...Instruction manual for Evario hospital bed 125 22 22...