Instruction manual for Evario hospital bed
158
DC gate for supply input
Phenomenon
EMC basic
standard
Immunity test level
Professional healthcare facilities
Short, transient electrical
disturbances / bursts
IEC 61000-4-4
+/- 2 kV; 100 kHz repetition frequency
Surges: conductor to
conductor
IEC 61000-4-5
+/- 0.5 kV; +/- 1kV;
Surges: conductor to
earth
IEC 61000-4-5
+/- 0.5 kV; +/- 1kV; +/- 2 kV;
Conducted disturbances
induced by high-
frequency fields
IEC 61000-4-6
3 V; 0.15 MHz to 80 MHz;
6V in ISM frequency bands between 0.15 MHz and
80MHz
80% AM at 1 kHz
Gates of patient connections
Phenomenon
EMC basic
standard
Immunity test level
Professional healthcare facilities
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 8 kV; contact
+/- 2 kV, +/- 4 kV ¸ +/- 8 kV, +/- 15 kV; +/- 25kV air
Conducted disturbances
induced by high-
frequency fields
IEC 61000-4-6
3 V; 0.15 MHz to 80 MHz;
6V in ISM frequency bands between 0.15 MHz and
80MHz
80% AM at 1 kHz
Gates of signal input/signal output parts
Phenomenon
EMC basic
standard
Immunity test level
Professional healthcare facilities
Electrostatic discharge
(ESD)
IEC 61000-4-2
+/- 8 kV; contact
+/- 2 kV, +/- 4 kV ¸ +/- 8 kV, +/- 15 kV; +/- 25kV air
Short, transient electrical
disturbances / bursts
IEC 61000-4-4
+/- 1 kV; 100 kHz repetition frequency
Conducted disturbances
induced by high-
frequency fields
IEC 61000-4-6
3 V; 0.15 MHz to 80 MHz;
6V in ISM frequency bands between 0.15 MHz and
80MHz
80% AM at 1 kHz
Содержание Evario
Страница 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Страница 125: ...Instruction manual for Evario hospital bed 125 22 22...
Страница 166: ...Instruction manual for Evario hospital bed 166 Notes...
Страница 167: ...Instruction manual for Evario hospital bed 167...
Страница 168: ...Australian Distributor...