Instruction manual for Evario hospital bed
144
Problem
Possible causes
Rectification
Manual CPR release of
backrest is not possible
Bowden cables are too loose or
not attached
Readjust at the release lever or
secure them
Bowden cable is kinked
Insert new Bowden cables; in-
form your operator about any
necessary repairs
Mains control lamp in control
unit does not light up
No power supply to socket
Use an electric socket that is
working properly
Mains cable damaged
Replace the mains cable
Fuses in control unit defective
Exchange the control unit; inform
your operator about any neces-
sary repairs
Actuator runs for a brief time
only, then stops; can then only
be activated in the opposite
direction.
Actuator overloaded
Remove the overload (run actua-
tor in in the opposite direction for
a brief time), retest
One or more motors are not
connected
Connect all motors
Control unit is not functioning:
Display screen
If available:
locking LEDs on locking
box/control box
There is a problem with the con-
trol unit. For safety reasons, all
functions are locked.
Troubleshoot and resolve (see
Chapter 15.2 to 15.5), then:
RESET control unit:preferably
via control box / locking box
(see Chapter 15.6); alternatively
also via LCD handset / control
panel (see Chapter 15.3)
If problem recurs: Have drive
system checked. Inform your
operator about any necessary
repairs.
Height adjustment and tilting
not functioning; acoustic signal
during adjustment
Control
unit has “forgotten” the
actuator positions
Initialise the control unit; (see
Chapter 14.9)
Tab. 5: Faults and their Rectification
Содержание Evario
Страница 26: ...Instruction manual for Evario hospital bed 26 Fig 3 Positions for storing the LCD handset...
Страница 125: ...Instruction manual for Evario hospital bed 125 22 22...
Страница 166: ...Instruction manual for Evario hospital bed 166 Notes...
Страница 167: ...Instruction manual for Evario hospital bed 167...
Страница 168: ...Australian Distributor...