TIP STRIP FAILURE DETECTION
RoboSep™ has an automatic tip strip failure detector. If the unit detects a tip strip failure, a dialog box will open (Figure 37) and the
machine will beep. Pressing the “Resume” button will allow RoboSep™ to try stripping the tip again. Following a successful tip strip, a
second dialog box will open (Figure 38). If the tip strip is still not successful, the tip can be removed manually. If a tip strip failure
occurs, the tip head may need to be cleaned as described in Section 5.3.
5.6
REPLACING THE FUSES
The two fuses are located at the bottom right side of the instrument, next to the power supply outlet and the power switch. If
RoboSep™ does not turn on when plugged in and the power switch is activated, the fuses may need to be replaced. To replace the
fuses:
1. Disconnect the power cable from the power source.
2. Disconnect the power cable from the instrument.
3. Open the fuse box and examine the fuses.
4. Replace the faulty fuse(s). Always use Type F 250V 4A 5X20mm fuses.
5. Reinsert the fuse box containing the new fuse(s) in its slot. Press gently until it snaps into place.
6. Reconnect the power cable to the instrument, then to a grounded (three-prong) power source.
Figure 37 – Dialog box shown if a tip strip failure is
detected
Figure 38 – Dialog box shown following tip strip failure
recovery