CG635 Remote Programming
37
CG635 Synthesized Clock Generator
3. Q / Q
¯ high and low levels and step sizes.
4. CMOS high and low levels and step sizes.
5. Current display selection.
6. PRBS on/off state
7. Run/stop state
8. Stop level
Example
*SAV 3<CR>
Save current settings to location 3.
*TST? Self
Test
Runs the instrument self test and returns 0 if successful and 72
(EXE_FAIL_SELF_TEST) if unsuccessful. If unsuccessful, the error buffer will
include device dependent errors related to the parts of the self test that failed.
Example
*TST?<CR>
Run the instrument self test and return the result.
*WAI
Wait for Command Execution
The instrument will not process further commands until all prior commands
including this one have completed.
Example
*WAI<CR>