ST535C Handbook
Page
31
Setup Parameters (0
-
31)
No.
Function
Min.
Max.
Default
Notes
0
Thermocouple type
0
3
2
0=K, 1=N, 2=R, 3=S
1
Error 1 enable
0
1
1
0=disabled, 1=enabled
2
Max. user temperature
100
1550
1320
°C
3
Display brightness
0
15
7
0=dim, 15=bright
4
Error 4 enable
0
1
1
0=disabled, 1=enabled
5
Error 5 enable
0
1
0
0=disabled, 1=enabled
6
Error 6 firing hours trip
10
1000
1000
1000=disabled
7
Room temperature trip
30
71
50
°C. 71=disabled
8
Power fail recovery enable
0
1
1
0=disabled,1=enabled
9
Paused time limit (hours)
1
11
2
11=disabled
10
Set point offset (zone 1)
-
99
99
0
°C
11
Proportional band (zone 1)
1
999
55
°C
12
Integral time (zone 1)
0
9999
200
Seconds, 0=disabled
13
Differential time (zone 1)
0
999
10
Seconds, 0=disabled
14
Kiln element power (zone 1)
0
9999
0
1 unit = 0.1kW
19
Error 9 differential temperature trip
0
99
0
°C. 0=disabled
20
Set point offset (zone 2)
-
99
99
0
°C
21
Proportional band (zone 2)
1
999
5
°C
22
Integral time (zone 2)
0
9999
0
Seconds, 0=disabled
23
Differential time (zone 2)
0
999
0
Seconds, 0=disabled
24
Kiln element power (zone 2)
0
9999
0
1 unit = 0.1kW
30
Set point offset (zone 3)
-
99
99
0
°C
31
Proportional band (zone 3)
1
999
5
°C
Note: The setup parameters shown thus are freely adjustable. The other setup
parameters might be password protected
-
contact supplier.
Содержание ST535C
Страница 34: ...Page ST535C Handbook 34 Notes ...