Meter Setup
Power Line Freq
[60 Hz]
Power Line Freq provides measuring immunity to industrial noise from a power supply feed.
To set Power Line Frequency, follow these steps from the
Meter Setup
menu:
1. Select
Power Line Freq
to view the
Power Line Frequency
display.
2. Position the arrow next to
50 Hz
or
60 Hz
.
3. Press
[E]
to save the power line frequency and return to the
Meter Setup
menu.
Excitation Freq
[Factory Set]
The Excitation Freq parameter is set at the factory. You can change Excitation Freq to configure the DC
excitation of the coils. Supported frequencies are dependent on the configured power line frequency:
50 Hz
60 Hz
1 Hz
1 Hz
3.125 Hz
3.75 Hz
6.25 Hz
7.5 Hz
12.5 Hz
15 Hz
To change Excitation Frequency, follow these steps from the
Meter Setup
menu:
1. Select
Excitation Freq
to view the
Excitation Frequency
display.
2. Position the arrow next the desired frequency.
3. Press
[E]
to save the excitation frequency and return to the
Meter Setup
menu.
Pipe Diameter
[Factory Set]
The Pipe Diameter parameter is set at the factory. In the event the amplifier is replaced, verify that the
pipe diameter matches the installed pipe size.
To change Pipe Diameter, follow these steps from the
Meter Setup
menu:
1. Select
Pipe Diameter
to open the
Pipe Diameter
display.
2. Position the arrow next to one of the following pipe diameters.
6 mm [1/4"]
65 mm [2-1/2"]
400 mm [16"]
1000 mm [40"]
8 mm [1/4"]
80 mm [3"]
450 mm [18"]
1050 mm [42"]
10 mm [1/4"]
100 mm [4"]
500 mm [20"]
1200 mm [48"]
15 mm [1/4"]
125 mm [5"]
550 mm [22"]
1400 mm [54"]
20 mm [1/4"]
150 mm [6"]
600 mm [24"]
1600 mm [64"]
25 mm [1"]
200 mm [8"]
700 mm [28"]
1800 mm [72"]
32 mm [1-1/4"]
250 mm [10"]
750 mm [30"]
2000 mm [80"]
40 mm [1-1/2"]
300 mm [14"]
800 mm [32"]
Other
50 mm [2"]
350 mm [14"]
900 mm [36"]
—
3. Press
[E]
to save the pipe diameter and return to the
Meter Setup
menu.
SSK
513
Electromagnetic Flow Meter
Page 30
March 2014