
13
Distance Calibration
Calibrate the distance readings of the walking profiler by measuring out 528 ft (160 meters) with a
rolling wheel measuring device. The calibration track should be as straight and level as possible. Once
the test track is prepared, start the calibration procedures through the Calibrate icon in the Collect
window. Select Distance Calibration and follow the steps precisely to complete a successful
calibration.
Set the length of the calibration track. The units can be changed by clicking on the feet (ft) symbol and
choosing the appropriate units. After the length of the track has been entered, select accept.
Once the walking profiler’s front wheel is on the beginning of the track, select
next and push the
device until the front wheel is at the end of the calibration track as indicated in the collection window.
Figure 9: The initial window of the distance calibration for setting the track legth.
Figure 10: The instructions for the distance calibration are shown above.
Содержание CS8900
Страница 30: ...29 Figure 32 A collection without events selected Figure 33 A collection with a point event at 0 14 3 feet...
Страница 39: ...38 Figure 41 The collection data included in a sidewalk report Figure 40 The top of a sidewalk report...
Страница 47: ...46 Figure 53 Selecting a location to save the exported file Figure 54 The export type drop down menu...