Operation
14
CTC100 Programmable Temperature Controller
Turn the instrument on
Plug the CTC100 in and turn it on with the power switch on the back of the instrument.
If the CTC100 does not turn on, a fuse may have blown. Two internal fuses can be accessed by
unplugging the instrument and then removing its top cover.
If the instrument boots up but does not respond to button or touchscreen presses, the most likely
cause is that the ground prong of the electrical socket is not connected to ground. A three-prong
plug must be used and all three prongs must be connected.
The Select screen
, arranged in roughly the same order as the
connectors on the back panel.
The AIO column shows the four analog I/O channels, while the DIO column shows the digital
I/O channel, the alarm relays, and the three virtual channels (V1, V2, and V3) that can be used to
perform real-time calculations.
The Select screen controls which channels appear on the Numeric, Plot, and Channel Setup
screens. To select a channel, touch a button on the Select screen; the button becomes lighter,
indicating that the channel is selected. Touch the button again to deselect the channel.
C onfigure the sensor inputs
Select the sensor type and calibration curve as follows:
1.
.
key. The top of the screen has one tab for each selected channel. Touch
one of the tabs to display the settings for that channel.
3.
etc.)
.
Содержание CTC100
Страница 1: ...Version 2 1 May 14 2019 User Manual CTC100 Cryogenic Temperature Controller...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 25: ...Operation 13 CTC100 Programmable Temperature Controller...
Страница 85: ...Operation 73 CTC100 Programmable Temperature Controller To disable cascade cont select it...
Страница 128: ......
Страница 129: ...Remote Programming 117 CTC100 Programmable Temperature Controller...
Страница 130: ......
Страница 142: ......
Страница 150: ......
Страница 176: ......