Remote Programming
105
CTC100 Programmable Temperature Controller
resumes if it was running to begin with. This feature can protect equipment from the excessive
temperatures that can occur if a PID feedback loop is poorly tuned.
To turn this feature off, issue the
alarm.output
command with an empty argument, i.e.:
In1.alarm.output()
Channel.
alarm.relay { None, A, B, C, D }
An alarm can trigger one of
relays. The
alarm.relay
instruction
determines which, if any, of the four relays is triggered.
Channel.
alarm.sound { None, 1 beep, 2 beeps, 3 beeps, 4 beeps }
Controls which sound plays when the alarm goes off.
C hannel.cal submenu
All
channel.cal
instructions are only available for input channels.
Channel.
cal.A
Channel.
cal.B
Channel.
cal.C
Channel.
cal.R0
0.0
These instructions set custom calibration coefficients for RTD, thermistor, or diode inputs with
a custom calibration type. See the description of the A, B, C, and R0 buttons on page 64 for more
information.
Errors: If
cal.Type
is not set to
Custom
, attempting to set
cal.A
,
cal.B
, or
cal.C
produces a run-
channel that is not a sensor input produces an assembly-
Channel.
cal.Gain
0.0
Channel.
cal.Offset
0.0
on. The offset and gain are applied
after the sensor signal is converted to temperature. These instructions provide an easy way to
make adjust
Errors: Attempting to set the offset or gain on a channel that is not an input produces an
assembly-
Channel.
cal.Type { IEC751, US, Custom }
(RTD sensor type)
Channel.
cal.Type { 100, 300, 1000, 2252, 3000, 5000, 6000, 10000B,
10000H, 30k, 100k, 300k, 1M, Custom }
(Thermistor sensor type)
Channel.
cal.Type { DT-470, DT-670, Si410, Si430, Si440, S700, S800,
S900, Custom }
(Diode sensor type)
Channel.
cal.Type { RX-102A, RX-103A, RX-202A, RO600, R400, R500 }
(ROX
sensor type)
Channel.
cal.Type { Custom, Standard }
(Channels with custom calibration tables)
Determines which calibration curve is used for a particular channel. The available arguments
depend on the value of the
channel
.Sensor
setting. See the description of the Type button on
page 63 for more information.
C hannel.P ID submenu
All
channel.
PID
instructions can only be used on output channels. Attempting to use a PID
instruction on an input cha
Содержание CTC100
Страница 1: ...Version 2 1 May 14 2019 User Manual CTC100 Cryogenic Temperature Controller...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 25: ...Operation 13 CTC100 Programmable Temperature Controller...
Страница 85: ...Operation 73 CTC100 Programmable Temperature Controller To disable cascade cont select it...
Страница 128: ......
Страница 129: ...Remote Programming 117 CTC100 Programmable Temperature Controller...
Страница 130: ......
Страница 142: ......
Страница 150: ......
Страница 176: ......