Remote Programming
98
CTC100 Programmable Temperature Controller
system.log.clear { yes, no }
Erases log files from the USB device.
system.log.clear USB
erases all logged data from internal memory and is mainly used
to remove old data from the plot screen.
system.log.clear RAM
erases all log files from the current log folder on the USB
device.
system.log.clear all
clears logged data from both USB and RAM.
system.log.clear?
always returns
no
.
system.log.folder "
Folder name
"
Determines which folder on the USB memory device
logged data. If the
folder does not exist, it is created. If the folder does exist and it already contains CTC100 logfiles,
new data points are appended to the existing files.
system.log.interval { off, "0.1 s", "0.3 s", "1 s", "3 s", "10 s", "30
s", "1 min", "3 min", "10 min", "30 min", "1 hr" }
log. Individual channels can override this value using the
channel.
logging
instruction.
system.log.Log to { RAM, USB, None }
Set this parameter to USB to begin logging data to a USB memory device, if one is present. Set
data appears on the Plot screen.
Errors
RAM
system.log.USB { Auto, Manual }
If set to
Auto
, any time a memory device is plugged into one of the CTC100 s USB ports, the
CTC automatically begins logging to it. If set to
Manual
, the user must touch the USB logging
triangle in the upper-right corner of the screen to begin logging.
system.other.A/D rate { 20 ms, 40 ms, 60 ms, 80 ms, 100 ms, 120 ms, 140
ms, 160 ms, 180 ms, 200 ms, 220 ms, 240 ms, 260 ms, 280 ms, 300 ms,
400 ms, 500 ms, 600 ms, 700 ms, 800 ms, 900 ms, 1000 ms }
(50 Hz line
frequency)
system.other.A/D rate { 16.7 ms, 33.3, 50 ms, 66.7 ms, 83.3 ms, 100 ms,
150 ms, 200 ms, 250 ms, 350 ms, 400 ms, 500 ms, 600 ms, 700 ms, 800
ms, 900 ms, 1000 ms }
(60 Hz line frequency)
system.other.A/D rate 0.0
(1 MHz trigger source)
Sets the A/D conversion time and determines how often PID feedback loops run. Different
arguments are available depending on whether the line frequency is 50 or 60 Hz. If
on the CTC100
the A/D
sampling can set to any value between 10 and 1000 ms.
system.other.fan { off, low, medium, high, max, auto }
Controls the system fan speed. Settings other than
auto
can reduce the accuracy of temperature
measurements and cause the DC outputs to overheat and shut down.
system.other.date "
month day year
"
system.other.time "
hours:minutes
"
Sets the time and date. Note that setting the time and date can adversely affect the display of
previously-
date string
should include the month, day, and year
Содержание CTC100
Страница 1: ...Version 2 1 May 14 2019 User Manual CTC100 Cryogenic Temperature Controller...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 25: ...Operation 13 CTC100 Programmable Temperature Controller...
Страница 85: ...Operation 73 CTC100 Programmable Temperature Controller To disable cascade cont select it...
Страница 128: ......
Страница 129: ...Remote Programming 117 CTC100 Programmable Temperature Controller...
Страница 130: ......
Страница 142: ......
Страница 150: ......
Страница 176: ......