Remote Programming
97
CTC100 Programmable Temperature Controller
"system.display.Monitors" { Show, Hide, Off }
If set to Show, the temperature of each I/O card that has a printed circuit board (PCB)
temperature sensor is collected, stored, and shown on the Select screen. The system must be
restarted before any changes are effective.
system.display.Type { Single, Multiple, Custom, Ponytail }
Controls the type of plot. On a Single plot, all selected channels appear on a single Y axis.
selected channel.
ffset by its initial value. The offset is recalculated
whenever the user scrolls or zooms the graph.
system.display.Units { °C, K, mK, °F, "" }
Sets the system units. Setting the units does not change previously-acquired data; that is, if a
value of 22°C is recorded in the log and the units are then changed to °F, graphs and logs will
appear to show that a value of 22°F was recorded. If the units are set to an empty string,
thermocouple readings are shown in millivolts and RTD and thermistor readings are shown in
ohms, even if the sensors have a custom calibration c
system.display.XLabels { None, Absolute, Elapsed }
Controls the type of label shown at the bottom of the plot. Absolute shows the time and date,
while Elapsed shows the relative time in seconds, minutes, hours, or days.
system.display.XRange
0
Sets the X range of the plot in milliseconds. Only the plot for the currently-selected group is
affected.
Errors: a run-time error occurs if the argument is less than 10000 (10 seconds) or greater than
2592000000 (30 days).
system.IP.Address
0.0.0.0
Sets the CTC100's IP address. The IP address should be in dotted-decimal notation, i.e.
"172.16.0.0".
Errors: If part of the specified IP address is not in the correct format (i.e. contains a non-
numeric character or a value that is not between 0 and 255), that portion of the IP address is set to
zero. The IP address cannot be changed if
system.IP.DHCP
is set to
on
.
system.IP.DHCP { On, Off }
Enables or disables the Dynamic Host Configuration Protocol. If DHCP is enabled and a DHCP
server is available on your network, the IP address is automatically set and cannot be changed
manually.
system.IP.MAC
00:00:00:00:00:00
Sets or queries the media access control address. This value is set at the factory and should not
generally be changed. The address should be specified in six groups of two hexadecimal digits
separated by colons, i.e., "00:19:b3:06:00:00"
The default MAC address is 00:19:b3:06:ab:cd, where abcd is the last three digits of the
instrument's serial number in hexadecimal format.
system.IP.Port
23
Sets the telnet port for Ethernet communications. Remote commands can be sent to the CTC
via a telnet connection on the selected port. The port must be a value between 0 and 65535,
inclusive, and should normally be set to either 23 (the default) or a value greater than 1024.
Содержание CTC100
Страница 1: ...Version 2 1 May 14 2019 User Manual CTC100 Cryogenic Temperature Controller...
Страница 6: ......
Страница 8: ......
Страница 12: ......
Страница 25: ...Operation 13 CTC100 Programmable Temperature Controller...
Страница 85: ...Operation 73 CTC100 Programmable Temperature Controller To disable cascade cont select it...
Страница 128: ......
Страница 129: ...Remote Programming 117 CTC100 Programmable Temperature Controller...
Страница 130: ......
Страница 142: ......
Страница 150: ......
Страница 176: ......