CALIBRATION (Cont’d) 

 
 
 
 
 
 
 
 
 
 

SETTING CALIBRATION 

 

NOTE:

  Ensure that nothing is in contact with the scale system during this procedure.  Remove 

hands from the system when noting the displayed calibration results.   
 

STEP 1:

  Remove  the  four  (4)  screws  on  the  bottom  of  the  display  housing.    Lift  the  display 

panel.  Put the scale system into the Calibration Mode by switching the calibration switch on the 
display board (Figure 9).  “

CALIBRATION

” will flash on the display.   

STEP 2: 

Select known calibrated weights, traceable to NIST. 

NOTE:

  The weights should total at least 227kilograms (half of the maximum scale capacity).  Put 

aside a subset of weight that is less than half of the combined weight.  

DO NOT USE

 barbells or 

un-calibrated weights. 

STEP 3:

  Press the 

MENU

 button until “

FULL = 453.59 Kg

” is displayed.  Change “

FULL”

 to the 

value of the total weight of the selected calibrated weights.  Use the 

UP

 arrow button to select the digit 

and the 

DOWN

 arrow button to change digit.  Press 

ENTER

 to save changes.   

STEP 4:

  Press the 

MENU

 button until “

HALF = 226.79.00 Kg

” is displayed.  

Change

 “

HALF”

 

to the value of the subset of the selected calibrated weights.  Use the 

UP

 arrow button to select the 

digit and the 

DOWN

 arrow button to change the digit. 

 Press 

ENTER

 to save changes.   

STEP 5:

  Press 

MENU

 button until “

3 Pt Calibration

” is displayed.  Press the 

UP

 arrow button.   

STEP 6: 

Zero the scale by removing all weight from the platform.  Press the 

UP 

arrow button.   

STEP 7:

  Place the 

HALF

 weight from Step 4 on platform.  Allow weight to stabilize.  Press 

UP

 

arrow to save change. 

STEP 8:

  Place the 

FULL

 weight from Step 3 on platform.  Allow weight to stabilize.  Press 

UP 

arrow to save change. 

STEP 9

:  Press 

ENTER

 to save the calibration, or 

WEIGH 

to exit without saving

.  

 

STEP 10:

  Remove weight and switch the scale system out of the Calibration Mode on the display 

board (Figure 9).   

STEP 11:

  Place the display panel into the display housing.  Align the display panel with the four 

holes in the display housing.  Secure display panel with the four (4) mounting screws.   

CAUTION 

The  integrated  circuits  and  semiconductors  on 
the  printed  circuit  boards  may  be  damaged 
by  electrostatic  discharge  (ESD).    Be  sure  to 
use proper handling precautions at all times.  

Содержание SR Scales SR585i-KG

Страница 1: ...CROBAN_201006 Page 1 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by...

Страница 2: ...ANTIMICROBIAL PROTECTION 8 STORAGE AND TRANSPORTATION 8 CLEANING AND DISINFECTING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BASIC SYSTEM OPERATION 12 BATTERY REPLACEMENT 14 THEORY OF OPERATION 15 CALIBR...

Страница 3: ...n Scale DESCRIPTION QUANTITY Box 1 BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY 1 ea DISPLAY UNIT WITH CABLE 1 ea HANDRAIL SET WITH HARDWARE 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea MEDICAL GRADE...

Страница 4: ...splay cable through cable hole on the right STEP 3 Figure 1 Insert the four 4 3 8 16 screws 4 into the holes in Mast Assembly and Base Assembly as indicated Secure the Mast Assembly by tightening the...

Страница 5: ...le to the upright position Figure 2 Display Cable Connector and Battery Compartment STEP 10 Figure 3 Insert the Left 11 and Right 12 Side Hand Rails into the Rail Holder cups on the sides of the Base...

Страница 6: ...Bracket 9 to the rear of the Mast Assembly 1 using the two 2 mounting screws provided NOTE The three holes on the Printer Mast Bracket should be towards the top of the Mast Assembly figure 4 STEP 2 Ap...

Страница 7: ...flat head Mast Base Assembly mounting screws Item 4 Figure 1 to allow movement of the Mast for insertion of the Height Bar Assembly into the Base Assembly STEP 2 After Height Bar Assembly installation...

Страница 8: ...weight is displayed on a 16 character dot matrix LCD The weight data may be viewed in kilograms with a displayed resolution of 0 1 kg INTENDED USE The SR585i KG Bariatric Stand On Scale is specifical...

Страница 9: ...ct areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available...

Страница 10: ...ange AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds RECALL Recalls last stored weight height and BMI AVERAGING Automatic digital filter POWER SUPPLY...

Страница 11: ...indicator flashes then remains solid when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and...

Страница 12: ...en PRINT INVERTED and PRINT NORMAL Press ENTER to save changes UP Press UP to adjust height up from the default to increase the scale s on time or to select a digit when setting time and date DOWN Pre...

Страница 13: ...ing a cane for support on the scale place the cane on the scale while zeroing the system This will ensure that the patient s NET weight will be displayed CALCULATING BMI Press to calculate BMI When th...

Страница 14: ...p the scale on its side and unscrew the four 4 screws in the Battery Compartment Cover located on the base of the unit between the wheels STEP 3 Remove and replace ALL six 6 D cell batteries Refer to...

Страница 15: ...ometers serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unlo...

Страница 16: ...ace the subset of calibrated weight on the scale Wait for scale to stabilize note scale reading STEP 4 Place all of the calibrated weight on scale Wait for scale to stabilize note scale reading Remove...

Страница 17: ...ed weights Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 226 79 00 Kg is displayed Change HA...

Страница 18: ...system out of the Calibration Mode on the display board Figure 9 STEP 6 Place the display panel into the display housing Align the display panel with the four holes in the display housing Secure disp...

Страница 19: ...battery Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label...

Страница 20: ...or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles...

Страница 21: ...ng and Service Manual Part No MAN585i KG_MICROBAN_201006 Page 21 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrumen...

Страница 22: ...o MAN585i KG_MICROBAN_201006 Page 22 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstr...

Отзывы: