CONFIGURING SEND BUTTON
STEP 1:
Remove the four (4) screws on the bottom of the display housing. Lift the display
panel. Put the scale system into the Calibration Mode by switching the calibration switch on the
display board (Figure 9). “
CALIBRATION
” will flash on the display.
STEP 2:
Press the “
MENU
” key until “
DATA OUT = PRINT
” is displayed
STEP 3:
Press “
UP
” or “
DOWN
” key to select between “
EHR
” and “
”
STEP 4:
Press the “
ENTER
” key to save
STEP 5:
Switch the scale system out of the Calibration Mode on the display board (Figure 9).
STEP 6:
Place the display panel into the display housing. Align the display panel with the four
holes in the display housing. Secure display panel with the four (4) mounting screws.
EHR OUTPUT
Electronic Medical Records (EMR) / Electronic Health Records (EHR) technology is supported by
the i-series scales from SR Instruments. The EMR/ EHR software is not included with scale.
There are many different EMR/ EHR software available and the connectivity of the scale to the
software requires the services of a professional.
A null modem cable is required to communicate with the scale.
Scale output when send button is configured to Electronic Health Records is as follows:
<esc>R<esc>Wnnn.n<esc>Hmmm.m<esc>Bkk.k<esc>Uuu<esc>E
Where
:
R is read
W is weight
nnn.n is the weight in kg
H is height
mmm.m is the height in cm.
B is BMI
kk.k is the BMI
U is units (kg)
uu is KG
E is end of packet.