background image

Model SR585

i

 Bariatric Stand-On Scale 

Operating and Service Manual 
Part No.  MAN585

i

-KG_MICROBAN_201006 

 

Page 14 of 22 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

BATTERY REPLACEMENT 

 
 
 
 
 
 
 
 
 
 
 

STEP 1: 

The display will read “

REPLACE BATTERIES

”. 

STEP 2: 

(Figure  7)  Tip  the  scale  on  its  side  and  unscrew  the  four  (4)  screws  in  the  Battery 

Compartment Cover located on the base of the unit, between the wheels.  

STEP 3: 

Remove  and  replace  ALL  six  (6)  “D”  cell  batteries.    Refer  to  Battery  Compartment 

Cover label for placement. 

STEP 4: 

Press the “

WEIGH

” button to confirm display is working

STEP 5: 

Tightly close the cover.  

STEP 6: 

Zero the system. 

 

 
 
 
 
 
 
 
 
 

 

 

 

 
 

Figure 7: Battery Compartment 

 

                                            

 

 

 

 

 

 

 

 

 

 

 

WARNING 

 

DO  NOT  remove  batteries  while  scale  is  on. 
Allow  scale  to  shut  down  naturally  thru 
timing  out.  This  prevents  possible  voltage 
drain  on  the  printed  circuit  board  coin 
battery. 

Содержание SR Scales SR585i-KG

Страница 1: ...CROBAN_201006 Page 1 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by...

Страница 2: ...ANTIMICROBIAL PROTECTION 8 STORAGE AND TRANSPORTATION 8 CLEANING AND DISINFECTING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BASIC SYSTEM OPERATION 12 BATTERY REPLACEMENT 14 THEORY OF OPERATION 15 CALIBR...

Страница 3: ...n Scale DESCRIPTION QUANTITY Box 1 BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY 1 ea DISPLAY UNIT WITH CABLE 1 ea HANDRAIL SET WITH HARDWARE 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea MEDICAL GRADE...

Страница 4: ...splay cable through cable hole on the right STEP 3 Figure 1 Insert the four 4 3 8 16 screws 4 into the holes in Mast Assembly and Base Assembly as indicated Secure the Mast Assembly by tightening the...

Страница 5: ...le to the upright position Figure 2 Display Cable Connector and Battery Compartment STEP 10 Figure 3 Insert the Left 11 and Right 12 Side Hand Rails into the Rail Holder cups on the sides of the Base...

Страница 6: ...Bracket 9 to the rear of the Mast Assembly 1 using the two 2 mounting screws provided NOTE The three holes on the Printer Mast Bracket should be towards the top of the Mast Assembly figure 4 STEP 2 Ap...

Страница 7: ...flat head Mast Base Assembly mounting screws Item 4 Figure 1 to allow movement of the Mast for insertion of the Height Bar Assembly into the Base Assembly STEP 2 After Height Bar Assembly installation...

Страница 8: ...weight is displayed on a 16 character dot matrix LCD The weight data may be viewed in kilograms with a displayed resolution of 0 1 kg INTENDED USE The SR585i KG Bariatric Stand On Scale is specifical...

Страница 9: ...ct areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available...

Страница 10: ...ange AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds RECALL Recalls last stored weight height and BMI AVERAGING Automatic digital filter POWER SUPPLY...

Страница 11: ...indicator flashes then remains solid when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and...

Страница 12: ...en PRINT INVERTED and PRINT NORMAL Press ENTER to save changes UP Press UP to adjust height up from the default to increase the scale s on time or to select a digit when setting time and date DOWN Pre...

Страница 13: ...ing a cane for support on the scale place the cane on the scale while zeroing the system This will ensure that the patient s NET weight will be displayed CALCULATING BMI Press to calculate BMI When th...

Страница 14: ...p the scale on its side and unscrew the four 4 screws in the Battery Compartment Cover located on the base of the unit between the wheels STEP 3 Remove and replace ALL six 6 D cell batteries Refer to...

Страница 15: ...ometers serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unlo...

Страница 16: ...ace the subset of calibrated weight on the scale Wait for scale to stabilize note scale reading STEP 4 Place all of the calibrated weight on scale Wait for scale to stabilize note scale reading Remove...

Страница 17: ...ed weights Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 226 79 00 Kg is displayed Change HA...

Страница 18: ...system out of the Calibration Mode on the display board Figure 9 STEP 6 Place the display panel into the display housing Align the display panel with the four holes in the display housing Secure disp...

Страница 19: ...battery Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label...

Страница 20: ...or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles...

Страница 21: ...ng and Service Manual Part No MAN585i KG_MICROBAN_201006 Page 21 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrumen...

Страница 22: ...o MAN585i KG_MICROBAN_201006 Page 22 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstr...

Отзывы: