IM-P505-02-US
Issue 1
72
UTM20 Series Ultrasonic Transit-time Flowmeters
Diagnostics
Submenus
Options/Descriptions
SYSTEM
(Requires
service level
passcode
or higher if
security is
enabled.)
TIME
HH:MM:SS
(24 hour clock)
Displays the time
DATE
YYYY-MM-DD
Displays the date
POWER ON TIME
In seconds
SIMULATION MODE
OFF
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
Flow Simulation provides output
and display simulation based on
a percentage of the full scale flow
Simulation will not accumulate the
totalizers The range of simulation
includes 0 to 100% of the full scale
flow
Use the Flow Simulation parameter
to set the range of simulation in
10% increments
To change the Flow Simulation,
from the
INPUTS/OUTPUTS menu:
1 Select FLOW SIMULATION to
view the
Flow Simulation display
2 Click RIGHT ARROW to
increment the percentage by 10, or
click DOWN ARROW to decrement
the percentage
by 10
PASSCODE LEVEL
READ ONLY
OPERATOR SERVICE
ADMIN RECOVERY
Defines the parameters, screens
and actions available to a user
confirmation screen
CLEAR HISTORY
CLEAR HISTORY
confirmation screen
Clears all alarms, warnings, errors
and informational messages from
the ALARM HISTORY buffer This
is typically done after startup or
maintenance on the flow system is
successfully completed
FACTORY RESET
FACTORY RESET
confirmation screen
Resets all parameters to the values
on the device when it was shipped
from the factory Any settings made
will be reset
REBOOT
REBOOT confirmation
screen
Reboots the device The UTM20
meter does not require this manual
REBOOT for any procedure,
but it may be useful for system
troubleshooting