XRBHR/XRBD SERIES MONOBLOCK® MANUAL
12
118185-001 REV B
X-RAY TUBE RE-SEASONING PROCESS
for
XRB80PN210HR
Parameters
Step 1 Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Step 10 Step11 Step12
Voltage (kV)
50
60
70
80
85
80
80
80
70
60
50
40
Current (mA)
0.35
0.65
0.90
1.2
1.5
1.8
2.2
2.6
3.0
3.5
4.2
5.2
For units with idle
time greater than
30 days
20 min 20 min 20 min 20 min 20 min 20 min 20 min 20 min 20 min
20 min
20 min
20 min
For units with idle
time of 14-30 days
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
5 min
For units with idle
time of 7-14 days
60 sec 60 sec 60 sec 60 sec 60 sec 60 sec 60 sec 60 sec 60 sec
60 sec
60 sec
60 sec
For units with idle
time of 1-7 days
30 sec 30 sec 30 sec 30 sec 30 sec 30 sec 30 sec 30 sec 30 sec
30 sec
30 sec
30 sec