101501-695 REV B
40
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
Power Limit
A Power Limit fault occurs the combination of KV and mA (kV multiplied by
mA) exceeds the maximum rated power of the Monoblock.
This fault will produce the following results:
Respond to a Report Fault Command with Fault #008
Shutdown the unit
Change the status of pin 7 of the analog interface
to ‘Active low’.
This is a latched fault.
Interlock
The fault occurs when the X-ray Interlock Enable Signal (+24Vdc) is
removed from the Monoblock.
This fault will produce the following results:
Respond to a Report Fault Command with Fault #009
Shutdown the unit
Change the status of pin 7 of the analog interface to
‘Active low’.
This is a latched fault.
Temperature Warning
The fault occurs when the oil temperature inside the Monoblock is greater
than 58
O
C.
This fault will produce the following results:
Toggle the Over Temp (OT) LED on/off
Respond to a Request Fault command with Fault #011
This is not a latched fault and will not shutdown the unit
This fault will only remain active for as long as the oil temperature inside
the Monoblock is greater than58
O
C.
Preventative Maintenance
The fault occurs when the Monoblock has been in the field for greater than
4 years or the High voltage on time is greater than 15,000 hours. When
this fault occurs, it is recommended that the Monoblock is returned to
Spellman High Voltage for preventative maintenance.
Respond to a Report Fault Command with Fault #043
This is not a latched fault and will not shutdown the unit.
Fault Clear
All faults can be cleared by executing a Reset Fault (CLR) command or
powering down the Monoblock.