XRBHR/XRBD SERIES MONOBLOCK® MANUAL
6
118185-001 REV B
Chapter 2
I
NSPECTION
&
I
NSTALLATION
nitial inspection and preliminary checkout procedures
are recommended. For safe operation, please follow the
step-by-step procedures described in Chapter 3,
Operating Systems.
2.1 Initial Inspection
Inspect the package exterior for evidence of damage due to
handling in transit. Notify the carrier and Spellman
immediately if damage is evident. Do not destroy or
remove any of the packing material used in a damaged
shipment.
After unpacking, inspect the Monoblock and control
assembly for any visible signs of damage.
The Monoblock and control assembly are paired together
as a system. Verify that the serial number located on the
Monoblock label is also reference on the control assembly
label and vice versa. It is important that the serial number
pair of Monoblock and control assembly are installed and
operated as a pair for normal operation. If for any reason
it becomes necessary to return the system both the
Monoblock and control assembly should be returned
together.
The Spellman XRBHR/XRBD MONOBLOCK
Series
of X-ray source and components are covered by warranty.
2.2 Installation
Reference the XRBHR data sheet SHV#128132-001
or
XRBD data sheet SHV#128146-001and outline drawing
SHV#
408644-001for dimensions.
The unit can be mounted in any position. Four, M8x1.25
threaded mounting holes are provided on the top and
bottom of the Monoblock®. Threaded M6x1 holes are
also provided for exit port collimation mounting.
The Control assemble has four, M5 threaded holes 10mm
deep.
Do not block the vent holes in the top cover as indicated
on the XRBHR data sheet SHV#128132-001 or
XRBD
data sheet SHV#128146-001.
I