101501-695 REV B
10
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
4.5.3 Request kV Setpoint
Description:
The host requests that the firmware report kV setpoint.
Syntax:
<STX>VSET;<CSUM><CR><LF>
Response:
<STX><ARG>;<CSUM><CR><LF>
Where:
<ARG> = the voltage in tenths of a kV (.1 kV), in ASCII format.
Example:
<STX>1000;<CSUM><CR><LF> is a voltage setpoint reading of
100.0 kV
.