118142-001 Rev B
Page 76 of 104
Request mA
monitor
61
None
-
Request Filament
Feedback
62
None
-
Request Filament
Limit
63
None
-
Request Filament
Pre-Heat
64
None
-
Request –15V
LVPS
65
None
-
Request Faults
68
None
-
Turn HV on/off
98
1 ASCII
0 or 1
Program
Local/Remote
Mode
99
1 ASCII
0 or 1
6.5 RESPONSE
OVERVIEW
The command responses will follow the same format as outlined above in
section 6.1. This list is comprised of Commands with complex responses
only. Commands using a simple response will use the <$> character
(ASCII 0x24) as a “Success” response or a single character error code.
These responses will be eight ASCII characters in length.
Response Name
<CMD>
Response
Request kV
Setpoint
14
10 ASCII
Request mA
Setpoint
15
10 ASCII
Request Filament
Limit Setpoint
16
10 ASCII
Request Filament
Pre-Heat Setpoint
17
10 ASCII
Request Analog
Monitor
Readbacks
19
23-50
ASCII
Request Total
Hours High
Voltage On
21
13 ASCII
Request Status
22
11 ASCII
Request DSP
Software Version
23
17 ASCII
Содержание DXM100 Series
Страница 11: ...DXM100 MANUAL 5 118147 001 Rev D Figure 2 1 Unit Dimensions ...
Страница 16: ...DXM100 MANUAL 10 118147 001 Rev D Figure 3 3 Local Programming Via External Voltage Source ...
Страница 17: ...DXM100 MANUAL 11 118147 001 Rev D Figure 3 4 Remote Monitoring ...
Страница 18: ...DXM100 MANUAL 12 118147 001 Rev D RELAY Figure 3 5 Enable Interlock Logic Control ...
Страница 37: ...118142 001 Rev B Page 13 of 104 Figure 9 Web Page 1 Contact Information ...