Troubleshooting
SPECTRO GENESIS FES 27 — 17.07.2019
— 49
Malfunction
Cause
Remedy
System information:
Wavelength calibration
failed.
Wrong sample or wrong
preflush time.
Use the correct sample. Check
the measurement parameters.
Repeat wavelength calibration
Error message:
Network connection
No communication
between the PC and the
instrument.
Spectrometer not switched
on.
Address incorrect or used
for a different purpose.
Data cable not plugged in
or defective.
Check main switch.
Enter the correct IP address
(online help). Only qualified IT
staff are permitted to enter the
address.
Check cable.