Transportation/Setting-up
— 22
SPECTRO GENESIS FES 27 — 17.07.2019
4.8
Exhaust air (to the outside)
Warning of toxic substances!
Vapors may be generated from the sample material. These may be toxic
and/or corrosive.
In these cases, provide protective measures, e.g. breathing apparatus
The sample material may be toxic and/or corrosive.
Ensure that the external exhaust system functions properly.
Exhaust air nozzles are located at the top of the instrument.
The corrosion-free exhaust chimneys must be connected with the laboratory
exhaust system using a non-flammable, flexible and corrosion-resistant hose (ID
100 mm). The fan must be resistant to chemicals and designed to operate at
temperatures up to 80°C (176°F). The exact position of the exhaust nozzles is
shown in the figure below.
All installation must be performed in a professional manner and in accordance with
applicable engineering standards in order to prevent risk to life and limb and avoid
damage to material property.
Fig. 9
Instrument
Extraction capacity at the nozzle
1 (rear)
Generator
2 (front)
Plasma torch box housing
SOP
≥ 250 m³/h; ≥ 147 ft³/min
100 – 140 m³/h; 59 – 82 ft³/min
1
2