Maintenance
SPECTRO GENESIS FES 27 — 17.07.2019
— 39
6.6
Cleaning the load coil
Warning! Hot surface!
The Plasma torch, plasma interface, and load coil may be very hot.
Let the components cool for 10–15 minutes before commencing work on
them.
The load coil may corrode because of the sample vapors.
In order to clean the coil, proceed as follows:
1.
Switch off the instrument.
2.
Remove the Plasma torch, see page 33.
3.
Clean the load coil with a small brush (e.g. a toothbrush) and a grinding paste
(e.g. toothpaste).
Do not bend the load coil.
4.
Remove the grinding paste completely from the load coil.
5.
Reinstall the components.
The load coil is cleaned.