background image

User Manual 

at the maximum tonnage load capability of the low tonnage gauge 
fitted. Higher tonnage loads than the low tonnage gauge maximum can 
still be obtained on the 15T or 25T press if they are to be applied, but 
then the pressure relief valve assembly will have to be readjusted to 
allow for a higher tonnage load to be hand pumped and contained on 
the system and therefore the low tonnage load gauge control valve 

MUST

 also be turned off. 

 
As an example for use of the 0 to 5 ton low tonnage gauge kit P/N 
GS15055 when fitted to an 

Atlas™ 15T Manual Hydraulic Press P/N 

GS15011 (which can be operated to apply up to a 15 ton maximum 
load), set the pressure relief valve accordingly such that a maximum 
load for 5 tons can be hand pumped on the press. This maximum load 
is indicated on both the standard 0 to 15 ton load gauge and 0 to 5 ton 
low tonnage load gauges that have been fitted. Any further pulls on the 
pump handle of the press results in an excess pressure being vented 
off at the pressure relief valve. In this way the pressure relief valve acts 
as a safety device to not only prevent overloading of and damage to 
the low tonnage gauge mechanism, but also to any sample or die 
assembly in the pressing area of the press itself. 
 
If a tonnage load higher than 5 tons is needed for application, switch 
off the low tonnage load gauge at its control valve and re-adjust the 
pressure relief valve assembly for a higher tonnage load to be applied 
and contained. 
 
Therefore, use of an Atlas™ 15T Manual Hydraulic Press allows for 
any tonnage load up to 15 tons to be applied to a sample or an 
evacuable pellet die in the pressing area. However, if a finer reading of 
the load applied up to a maximum of 5 tons is required, then the low 
tonnage gauge kit P/N GS15055 can be fitted to the press and the 
additional low tonnage gauge provided is switched on-line. 
 
 
 
 

 

Содержание GS15051

Страница 1: ...Low Tonnage Gauge Kits for Atlas Manual Hydraulic Presses User Manual 2I 15051 8...

Страница 2: ......

Страница 3: ...Low Tonnage Gauge Kits for Atlas Manual Hydraulic Presses User Manual 2I 15051 8...

Страница 4: ...F THE ATLAS MANUAL HYDRAULIC PRESS WITH A LOW TONNAGE GAUGE FITTED 10 SAFETY WARNINGS WHEN OPERATING THE PRESS 10 BLEEDING AIR FROM THE LOW TONNAGE GAUGE KIT 10 SWITCHING ON AND OFF THE LOW TONNAGE GA...

Страница 5: ...from the pressure system but any pressure of oil in the system will always be registered as an appropriate applied tonnage load at the 15 or 25 ton gauge fitted The lower tonnage load gauge has finer...

Страница 6: ...and 0 to 5 ton low tonnage load gauges that have been fitted Any further pulls on the pump handle of the press results in an excess pressure being vented off at the pressure relief valve In this way t...

Страница 7: ...assembly with 0 to 1 0 to 2 or 0 to 5 tons load gauge fitted 1 Fixing bolt Fixing bolt with two Dowty seals is already in position at base of manifold body with a little red cap seal 1 Open ended span...

Страница 8: ...in the oil you may be able to pinpoint more readily what is needed to correct and to get the new system operational Procedure for Fitting To help in the explanation of the procedure for fitting a low...

Страница 9: ...F end of the spanner supplied Note The connection 39 nut at N1 must be turned clockwise using the 19mm A F spanner and holding the gauge 31 with your other hand to separate the threaded parts Fig 1 Lo...

Страница 10: ...ee Fig 1 and Fig 3 If the removed Dowty seal 48 from the press is not suitable for re use then use the spare seal provided with the low tonnage gauge kit Note The bore opening 76 on the manifold body...

Страница 11: ...Fitted to the Manifold Block Proceed to fit the complete manifold tube assembly with the two gauges via the fixing bolt 77 and Dowty seal 78 into position on the pump block assembly of the press from...

Страница 12: ...left open when higher tonnage loads are wished to be pumped on the press Bleeding Air from the Low Tonnage Gauge Kit Before the Atlas manual hydraulic press can be used to apply a tonnage load with t...

Страница 13: ...igher load tonnage gauge 31 fitted at the base of the manifold body 77 and towards the shut off valve tap assembly 79 Pump to apply a tonnage load of about 1 one ton no more against the metal block to...

Страница 14: ...apply if the LOWEST tonnage gauge of 0 to 1 tons range has been fitted to the manifold body 77 and this gauge is switched on line to register an oil pressure from the bleed of air procedure When the...

Страница 15: ...ee Fig 6 and the tap handle only needs a quarter to half turn maximum anti clockwise to open for the low tonnage gauge 75 to be switched on to register a tonnage load Beware If the valve tap 79 is ope...

Страница 16: ...e kit 75 Low tonnage gauge 0 to 1 0 to 2 or 0 to 5 ton range option 76 Bore opening on manifold body 77 Manifold body fixing bolt 78 Dowty seal 2 off for manifold body fixing bolt 79 Open close valve...

Страница 17: ...Notes for Use of Low Tonnage Gauge Kit...

Страница 18: ...Notes for Use of Low Tonnage Gauge Kit...

Страница 19: ...te www systems eng co jp Spain Teknokroma S Coop C Ltda Barcelona Tel 93 674 8800 Fax 93 675 2405 email comercial teknokroma es Switzerland Portmann InstrumentsAG Biel Benken Tel 061 726 6555 Fax 061...

Страница 20: ...PECAC INC 414 Commerce Drive Suite 175 Fort Washington PA 19034 USA Tel 215 793 4044 Fax 215 793 4011 SPECAC LTD Unit 12 Science Innovation Centre Halo Business Park Orpington Kent BR5 3FQ Tel 44 0 16...

Отзывы: