background image

 

Low Tonnage Gauge Kits for Atlas™ Manual Hydraulic Presses 

 

11 

that has been fitted, to stop any oil pressure pumped on the system 
from reaching it. The valve tap (

79

) is a tapered “needle” type fitting 

and will be closed when the tap handle (

79

) is turned hand tight in a 

clockwise direction. There is an O-ring (

80

) on the valve tap assembly 

that makes an oil-tight seal when fully closed. (See 

Fig 6

.)  

 

 

 
 

Fig 6. Low Tonnage Gauge Open/Close Valve Tap Assembly 

 
Close the pressure release handle (

30

) of the press by turning 

clockwise (see 

Fig 2.

) and begin to hand pump the press (slowly) in 

the usual way. Oil will be pressurized and forced from the press 
reservoir up and into the manifold body (

77

) to register at the higher 

load tonnage gauge (

31

) fitted at the base of the manifold body (

77

and towards the shut off valve tap assembly (

79

). Pump to apply a 

tonnage load of about 1 (one) ton (no more) against the metal block to 
be registered at the higher tonnage load gauge (

31

) as fitted. 

 
Now, very slowly open the valve tap (

79

) by anti-clockwise rotation of 

the handle until the O-ring (

80

) is just beginning to emerge from the top 

of the manifold body (

77

). (See 

Fig 7

.)  Any air that may be trapped in 

the manifold body (

77

) will be forced out at the top past the opened 

valve tap (

79

) and by the O-ring seal (

80

) from the oil pressure that has 

been pumped on the system. When air and oil is observed at the valve 
tap (

79

) by the O-ring (

80

), a slight decrease in the tonnage load that 

has been applied at the higher tonnage load gauge (

31

) and a 

registering of a low tonnage load at the low tonnage gauge (

75

) fitted 

will also be observed. 

80 

79 

Содержание GS15051

Страница 1: ...Low Tonnage Gauge Kits for Atlas Manual Hydraulic Presses User Manual 2I 15051 8...

Страница 2: ......

Страница 3: ...Low Tonnage Gauge Kits for Atlas Manual Hydraulic Presses User Manual 2I 15051 8...

Страница 4: ...F THE ATLAS MANUAL HYDRAULIC PRESS WITH A LOW TONNAGE GAUGE FITTED 10 SAFETY WARNINGS WHEN OPERATING THE PRESS 10 BLEEDING AIR FROM THE LOW TONNAGE GAUGE KIT 10 SWITCHING ON AND OFF THE LOW TONNAGE GA...

Страница 5: ...from the pressure system but any pressure of oil in the system will always be registered as an appropriate applied tonnage load at the 15 or 25 ton gauge fitted The lower tonnage load gauge has finer...

Страница 6: ...and 0 to 5 ton low tonnage load gauges that have been fitted Any further pulls on the pump handle of the press results in an excess pressure being vented off at the pressure relief valve In this way t...

Страница 7: ...assembly with 0 to 1 0 to 2 or 0 to 5 tons load gauge fitted 1 Fixing bolt Fixing bolt with two Dowty seals is already in position at base of manifold body with a little red cap seal 1 Open ended span...

Страница 8: ...in the oil you may be able to pinpoint more readily what is needed to correct and to get the new system operational Procedure for Fitting To help in the explanation of the procedure for fitting a low...

Страница 9: ...F end of the spanner supplied Note The connection 39 nut at N1 must be turned clockwise using the 19mm A F spanner and holding the gauge 31 with your other hand to separate the threaded parts Fig 1 Lo...

Страница 10: ...ee Fig 1 and Fig 3 If the removed Dowty seal 48 from the press is not suitable for re use then use the spare seal provided with the low tonnage gauge kit Note The bore opening 76 on the manifold body...

Страница 11: ...Fitted to the Manifold Block Proceed to fit the complete manifold tube assembly with the two gauges via the fixing bolt 77 and Dowty seal 78 into position on the pump block assembly of the press from...

Страница 12: ...left open when higher tonnage loads are wished to be pumped on the press Bleeding Air from the Low Tonnage Gauge Kit Before the Atlas manual hydraulic press can be used to apply a tonnage load with t...

Страница 13: ...igher load tonnage gauge 31 fitted at the base of the manifold body 77 and towards the shut off valve tap assembly 79 Pump to apply a tonnage load of about 1 one ton no more against the metal block to...

Страница 14: ...apply if the LOWEST tonnage gauge of 0 to 1 tons range has been fitted to the manifold body 77 and this gauge is switched on line to register an oil pressure from the bleed of air procedure When the...

Страница 15: ...ee Fig 6 and the tap handle only needs a quarter to half turn maximum anti clockwise to open for the low tonnage gauge 75 to be switched on to register a tonnage load Beware If the valve tap 79 is ope...

Страница 16: ...e kit 75 Low tonnage gauge 0 to 1 0 to 2 or 0 to 5 ton range option 76 Bore opening on manifold body 77 Manifold body fixing bolt 78 Dowty seal 2 off for manifold body fixing bolt 79 Open close valve...

Страница 17: ...Notes for Use of Low Tonnage Gauge Kit...

Страница 18: ...Notes for Use of Low Tonnage Gauge Kit...

Страница 19: ...te www systems eng co jp Spain Teknokroma S Coop C Ltda Barcelona Tel 93 674 8800 Fax 93 675 2405 email comercial teknokroma es Switzerland Portmann InstrumentsAG Biel Benken Tel 061 726 6555 Fax 061...

Страница 20: ...PECAC INC 414 Commerce Drive Suite 175 Fort Washington PA 19034 USA Tel 215 793 4044 Fax 215 793 4011 SPECAC LTD Unit 12 Science Innovation Centre Halo Business Park Orpington Kent BR5 3FQ Tel 44 0 16...

Отзывы: