![Spacelabs 91369 Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/spacelabs/91369/91369_service-manual_1339016051.webp)
91369 Service Manual
2-43
Setup
RESTART STATE STARTED/ENDED
Touch the RESTART STATE /STARTED key to set the default power ON state of the monitor to case mode.
When the monitor starts from a cold boot, it will start up with case mode enabled.
Touch the RESTART STATE /ENDED key to set the default power ON state of the monitor to non-case mode.
Refer to the
Ultraview SL Operations Manual
(P/N 070-1150-xx, located on CD-ROM P/N 084-1101-xx) for
how to use Start Case and End Case featues.
Screen Saver Setup
Touch the SCREEN SAVER SETUP key in the
Biomed Level
menu to select between the two screen saver
options (refer to
).
Figure 2-29: Screen Saver Setup dialog box
Reset Monitor
The RESET MONITOR key in the
Biomed Level
menu allows you to reset the monitor after changing settings
for the following items (the monitor must be reset before the changes can take effect):
• Monitor ID
• Monitor Name
• Subnet Name
• IP configurations
• Start Case/End Case options