91369 Service Manual
5-15
Troubleshooting
Troubleshooting Method 2: Using the Remote Terminal
1 Attach a remote terminal (9600 baud, no parity, 8 bits, 1 stop bit) to the serial connector.
2 Power ON the monitor.
3 Activate the
Serial Diagnostic
menu by pressing CTRL D during the “4, 3, 2, 1” countdown at system
startup.
4 All power-ON tests can be performed to isolate the problem to a PCBA. The monitor will guide you to the
appropriate field replaceable unit.
Touchscreen
If the display is present but the touchscreen is unresponsive:
1 Calibrate the touchscreen (refer to
on page 5-7) using the mouse. Check the operation
again.
2 Run the Interactive Touchscreen test (refer to
3 Remove the external power supply and battery(ies) and check the connections interfacing the touchscreen
to the CPU PCBA. If the connections are okay, replace the CPU PCBA.
Module Slot
1 Verify the module’s operation on another monitor (if available).
2 Check the software versions of all installed modules, Flexport interfaces, and other SDLC equipment
attached, verifying with Spacelabs Healthcare that there are no incompatibilities.
3 Inspect the Interconnect PCBA. If any connectors or parts appear damaged, replace the damaged
assembly.
4 If the problem persists, replace the CPU PCBA and the Interconnect PCBA, in that order.
SDLC Output
1 If a module housing is present, verify that the proper SDLC terminations are set and that the proper cables
are being used (refer to
for more information).
2 Check the software versions of all installed modules, Flexport interfaces, and other SDLC equipment
attached, verifying with Spacelabs Healthcare that there are no incompatibilities.
3 Inspect the I/O (Connector) PCBA and the Interconnect PCBA. If any connectors or parts on either of these
assemblies appear damaged, replace the damaged assembly.
4 If the problem persists, replace the CPU PCBA.