![Spacelabs 91369 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/spacelabs/91369/91369_service-manual_1339016019.webp)
91369 Service Manual
2-11
Setup
Network Installation
A typical network consists of bedside and central monitors and an optional clinical information system
(
Figure 2-9: Typical network configuration
Warning:
Ensure that the Ethernet wall plate and the shield of the Ethernet connecting cable are bonded to
the hospital grounding system.
Ethernet Network Connection
Caution:
• Only qualified personnel should attempt to connect a monitor to an Ethernet LAN the first time.
• Do not connect the monitor to an Ethernet local area network (LAN) prior to configuring the
following settings. The monitor must be properly configured for LAN access before you operate
the monitor. If you fail to correctly configure the monitor, you may interrupt other units also
using the LAN.
Note:
Detailed installation instructions for the physical Ethernet LAN are beyond the scope of this document.