Setup
91369 Service Manual
2-42
START/END CASE OPTIONS
Touch the START/END CASE OPTIONS key to display the
Start/End Case Options
dialog box.
Note:
If option D, Perioperative, is not activated, the START/END CASE OPTIONS key does not appear.
Figure 2-27:
Start/End Case
dialog box
Touch START/END CASE /ON to turn ON Start Case/End Case features, or touch START/END CASE /OFF to
turn the features OFF. The monitor must be reset for these changes to take effect, and the
Reset Monitor
dialog box (refer to
on page 2-44) displays after any change in the
Start/End Case Options
dialog
box.
The START/END CASE ON/OFF setting determines whether the following keys are enabled or disabled.
CASE PURGE QUERY/ALWAYS
The CASE PURGE /QUERY key determines whether the
Purge patient data?
confirmation window (refer to
on page 2-42) displays after the user touches KEEP SETTINGS on the
Keep Settings?
dialog
box.
Figure 2-28:
Purge patient data
confirmation window
Touch CASE PURGE /QUERY to prompt the user with the
Purge Data
confirmation window when the user
touches END CASE.
Touch CASE PURGE /ALWAYS to purge all patient data when the user touches END CASE.