![Spacelabs 91369 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/spacelabs/91369/91369_service-manual_1339016036.webp)
Setup
91369 Service Manual
2-28
Gateway Address xx.x.xxx.xxx is out of range or invalid
Refers to the
WLAN
tab. Monitors display this on the keyboard window’s error line or in a warning dialog box if
the form of the IP address is invalid (not in the form xxx.xxx.xxx.xxx, where X is a number) and SAVE is
selected.
WLAN IP address xxxxx is out of range or invalid
Refers to the
WLAN
tab. Same as for
Gateway Address xx.x.xxx.xxx is out of range or invalid
, where xxxxx is
the entered IP address.
WLAN Subnet Mask xx.x.xxx.xxx is out of range or invalid
Refers to the
WLAN
tab. Same as for
Gateway Address xx.x.xxx.xxx is out of range or invalid
Serial Ports
Refer to
on page 7-1 for the menu structure.
Patient Data Logger (Option R)
The Patient Data Logger option automatically sends patient vital signs from the monitor to a serial external
device, such as a printer or a terminal. Episodic patient data is also sampled and transmitted. The output is in
the form of ASCII text byte strings and is printed using standard RS-232 serial communications via the
monitor’s serial port (refer to the
Ultraview SL Operations Manual
,
P/N 070-1150-xx, located on CD-ROM
P/N 084-1101-xx for configuration information).
This option continues to send data whether the external device is on-line or off-line. Data transmission can be
stopped by reassigning the data port or disabling the Patient Data Logger option.
Communication between the monitor and the external device is set up by assigning the serial port to Patient
Data Logger and then adjusting the serial port settings. The various serial settings can be adjusted to suit the
device attached to the serial port.
To set up Patient Data Logger:
1 Touch MONITOR SETUP.
2 Touch PRIVILEGED ACCESS.
3 Enter the biomed password (default is
biomed
).
4 Touch SERIAL PORTS.
5 Touch ASSIGNMENT.
6 Touch DATA LOGGER.
7 Touch PREVIOUS MENU.