Respiration Monitoring
élance
Vital Signs Monitor Operations Manual
Page 10-3
2.
Touch
the
Alarms On
key
.
A dropdown list, as shown below, will appear.
3.
Touch
the
required Alarm
Selection.
4.
Touch
the
Low Limit
keys
to increase or decrease the value.
5.
Touch
the
High Limit
keys
to increase or decrease the value.
6.
Touch
the
OK key
to confirm.
10.4 Time to Respiration Alarm
Alarms
Time to Alarm
Time to Alarm
(without Alarm
(with Alarm
Validation)
Validation)
Respiration Rate - High Limit Violation Less than 14 seconds
Less than 24 seconds
Respiration Rate - Low Limit Violation
Less than 40 seconds
N.A.
Note
: In the event of interference caused by other devices in the care environment,
e tim to alarm may be slightly longer than 40 seconds.