Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-20
EtCO2 Conditions:
Condition
Display Value
Alarm Message
Priority
EtCO
2
within limits
<Number>
None
None
EtCO
2
< Low Limit
<Number>
EtCO
2
lower than Low Limit
Medium
EtCO
2
> High Limit
<Number>
EtCO
2
higher than High Limit
Medium
No Sampling line
connected
<Blank> None
(at startup / standby)
None
*Sampling line
disconnected
(during monitoring)
ank>
<Bl
CO2
unplugged
Low
Sampling tube is blocked ---
Occluded line
Low
Note:
*designates these technical alarms will be reset by touching the “Audio” icon.
ms will be turned off (including both visual and audio alarm
indicator) by touching the “Audio“ icon in the alarm state. The alarm will not be
r
m
or
6.10
ault P
r Ra
T
Signs
shipp
ettin
default
setting for
mode is Adult. Authorized personnel can use
res
d
ervice manual to change default settings.
Parameter
The technical alar
e-enabled until the alar condition is c rected.
Factory Def
aramete
nges and Settings
he
élance
Vital
Monitor is
ed with factory default s
gs. The
the patient
escribed in the s
the procedu
Ranges and Factory Default Settings
Setting Name
Default Value
Possible Values
HR high alarm limit
180 bpm
20-300 bpm
HR low alarm limit
50 bpm
15-295 bpm
HR alarms on
On
On
Pulse Tone Source
OFF
ECG, SpO
2
, OFF
RR high alarm limit
50 bpm
6-120 bpm
RR low alarm limit
6 bpm
5-119 bpm
RR alarms on
Off
On, Off
SpO
2
high alarm limit
100%
51-100%
SpO
2
low alarm limit
92%
50-99%
SpO
2
alarms on
Off
On, Off
PR alarms on
Off
On, Off
PR high alarm limit
180
35-240 BPM