SpO
2
Monitoring
élance
Vital Signs Monitor Operations Manual
Page 12-15
SatSeconds Indicator
When the SatSeconds algorithm detects an SpO
2
value outside the alarm limit,
the SatSeconds will begin to be counted, and the SatSeconds Indicator begins
to fill clockwise. Each second the saturation value is in violation of the
threshold is added to the SatSeconds count. When the SatSeconds count
meets or exceeds the SatSeconds setting, the SatSeconds indicator is
completely filled, and the SpO
2
alarm will be triggered.
When the saturation value is no longer in violation of the alarm limit, the alarm
will stop and the SatSeconds indicator begins to empty.
Note:
If an alarm threshold is crossed three or more times in a 60-second period, an
alarm will be triggered even if the SatSeconds limit has not been attained..
S
TEPS
:
1.
Touch
the
SpO
2
Numerical Area
.
A Setup SpO
2
window will appear.