Vital Signs Monitor Operations Manual
Page 12-17
4.
Select
the
required Response Mode
.
5.
Touch OK key
to confirm or
Cancel
key to quit.
When
No Pulse Alarm
is enabled, if a persistent loss-of-pulse condition is
detected, a medium level alarm will be triggered with
“Pulse Lost”
message
displayed in the message zone. When the
No Pulse
alarm condition is
indicated, always check on the patient.
S
TEP
window will appear.
2
O
2
window will
appear.
12.8.3.3
No Pulse Alarm Setting
S
:
1.
Touch
the
SpO
2
Numerical Area
.
A Setup SpO
2
2.
Touch
the
Advanced SpO Setting
key
.
A new Setup Sp
3.
Touch
the
No Pulse Alarm
key.
A dropdown list as shown below will
appear.
Se
4.
5.
lect
the
No Pulse Alarm
setting to turn it on of off.
Touch OK key
to confirm or
Cancel
key to quit.
12.8.3.4
Perfusion Index(PI)
Perfusion Index (PI) is a relative assessment of the pulse strength at the
measuring site. The system displays a "LOW PERFUSION" message