Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-15
HR Conditions:
Condition
Display
Value
Alarm Message
Priority
HR within limits
<Number>
None
None
HR < Low Limit
<Number>
HR lower than Low Limit
Medium
HR > High Limit
<Number>
HR higher than High Limit
Medium
HR Asystole
ASY
ECG Asystole
High
HR Ventricu
Fibrillation
lar
ular
Fibrillation
VF ECG
Ventric
High
Ventricular Tachycardia
TACH
chycard
High
V
ARR ventricular ta
ia
PVC/MINUTE PVC/Min
ARR
PVC/MINUTE Medium
Pause Pause
ARR
Pause Medium
Couplet
Couplet
ARR ventricular couplet
Medium
Pacer Non-capture
Pacer
ptu
PACEMAKER PACER
URE
non-ca
re
NON-CAPT
Medium
HR Lead-off
standby
lank>
None
(after start-up /
)
<B
None
†
(a
HR Lead-off
fter leads applied)
<Blank>
ECG Lead off
Low
HR Artifact
---
ECG Artifact
Low
ST within limits
one
<Number>
None
N
ST Lead # < Low Limit
r>
ST SEGMENT ALARM LEAD #
Medium
<Numbe
ST Lead # > High Limit
ber>
NT ALARM LEAD
Medium
<Num
ST SEGME
#
C
†
aution:
HR Lead-off technical alarm will be triggered only for the 1
st
ECG
lead on display when HR source is from ECG.