Managing Alarms
élance
Vital Signs Monitor Operations Manual
Page 6-22
Setting Name
Default Value
Possible Values
P2s/P2 ARTs/P2 PRSs
alarms on
Off On,
Off
P2m/P2 ARTm/P2 PRS
low alarm limit
m
70 mmHg
-10 to 130 mmHg
P2m/P2 ARTm/P2 PRSm
Hg
high alarm limit
130 mmHg
135 to 230 mm
P2m/P2 ARTm/P2 PRSm
alarms on
Off On,
Off
P2d/P2 ARTd/P2 PRSd low
alarm limit
60 mmHg
-10 to 120 mmHg
P2d/P2 ARTd/P2 PRSd high
Hg
alarm limit
120 mmHg
125 to 210 mm
P2d/P2 ARTd/P2 PRSd
alarms on
Off On,
Off
P2 CVPm/P2 LAPm/P2
RAPm low alarm limit
0 mmHg
-10 to 225 mmHg
P2 CVPm/P2 LAPm/P2
RAPm high alarm limit
20 mmHg
-5 to 230mmHg
P2 CVPm/P2 LAPm/P2
Off On,
Off
RAPm alarms on
TEMP high alarm limit (Note:
for
39.0 C
0 F
32.1-42.0 degrees C
.6 degrees F
depends on user setting
TEMP units)
102.
89.7-107
TEMP low alarm limit (Note
depends on user
:
setting for
97.0 F
89.6-107.5 degrees F
TEMP units)
36.0 C
32.0-41.9 degrees C
TEMP alarms on
Off
On, Off
EtCO
2
high alarm limit
60 mmHg
8 Kpa
mHg
0.8 to 10.6 Kpa
7 to 80m
EtCO
2
low alarm limit
20 mmHg
2 Kpa
0.6 to 7.9 Kpa
5 to 60 mmHg
EtCO
2
alarm on
Off
On, Off
FiCO
2
high alarm llimit
8 mmHg
1 Kpa
mHg
0.0 to 3.4 Kpa
0 to 24 m
FiCO
2
alarm on
Off
On, Off
RR(from EtCO
2
) high alarm
30 bpm
6 to 150 bpm
limit
RR(from EtCO
2
) low alarm
5 bpm
5 to 149 bpm
limit
RR(from EtCO
2
) alarm limit
Off On,
Off
on
ECG Waveform Size
10 mm/mV
2.5, 5, 10, 20 mm/mV
Resp Waveform Size
Auto
Auto