Cleaning
élance
Vital Signs Monitor Operations Manual
Page 18-3
ocol for the handling of blood and body fluids.
Signs Monitor.
18.2
cerning disposable and
infection,
n those specified by
tal Signs Monitor’s
18.3 Noninvasive Blood Pressure Cuffs
and disinfected with an
tion of household bleach (5.25% sodium
18.3.1
such as ENZOL (US) or CEDEZYME (UK)
10% solution of household bleach (5.25% sodium hypochlorite) in distilled water
s
Spray
bottles
tilled water.
. Spray bleach solution on the affected area until saturated. Allow the cuff to sit for
five minutes.
ion with a soft cloth and rinse again with distilled water.
ing at ambient temperature.
e hose connector.
Note:
Follow your hospital prot
Do not allow liquid to enter the
élance
Vital
Accessories
Where provided, follow the manufacturers' instructions con
reusable supplies.
As applicable, follow your hospital protocol concerning cleaning, dis
and/or sterilization of reusable supplies.
Use of patient cables, transducers, sensors, or supplies other tha
pacelabs Healthcare may adversely affect the
élance
Vi
S
performance.
r
The reusable cuff is packaged non-ste
enzymatic detergent and 10% solu
ile. It may be cleaned
hypochlorite).
Cuff Cleaning and Disinfection
Materials
Enzymatic detergent
Soft cloths or bristle brushe
S
TEPS
:
1. Prepare the enzymatic detergent and bleach solutions in separate bottles
according to the manufacturer’s instructions.
2. Spray detergent liberally on cuff, allowing it to sit for one minute.
. Remove detergent with a soft cloth. For persistent contamination, scrub with a soft
3
bristled brush.
4. Rinse the cuff thoroughly with dis
5
6. Remove any excess solut
Allow two hours for air dry
Note:
Make sure water does not enter th