3-3
070-2177-01 Rev B
Patient Safety
Failsafe machanisms limit the inflation period to less than 180s and the absolute maximum
pressure to 300mm Hg.
The air hose is detatchable by the patient.
Service Life
The service life of the OnTrak 90217A and associated accessories will be 7 years minimum.
Classification
EN 60601-1
Internally powered equipment.
Type BF-defibrillator proof applied part.
Not suitable for use with flammable gases.
Rated for intermittent operation.
Electrical Requirements
Power Requirement - Three AA size (LR6 or equivalent), alkaline or rechargeable batteries.
Physical Dimensions
Height
0.97 in (2.5 cm)
Depth
3.94 in (10.0 cm)
Width
2.77 in (7.0 cm)
Weight excluding batteries.
90217A
6.1oz (172g)
Environmental Requirements
Storage
Temperature
29 °F to 149 °F (-34 °C to 65 °C)
Humidity
95% (non-condensing)
Altitude
500 to 15,000 ft (-152 to 4,572 m)
Operating
Temperature
41°F to 104 °F (5 °C to 40 °C)
Humidity
90% (non-condensing)
Altitude
0 to 10,000 ft (0 to 3.48 m)
Atmospheric Pressure
700 hPa to 1060 hPa
Electromagnetic Compatibility
EN 60601-1-2: 2001, EN 60601-2-30
Emissions
CISPR11/FCC Part 15 Group 1 Class B
Содержание 90217A
Страница 4: ...ABP 90217A Operations Manual iv This page is intentionally blank...
Страница 24: ...ABP 90217A Operations Manual 1 18 This page is intentionally blank...
Страница 36: ...ABP 90217A Operations Manual 3 10 This page is intentionally blank...
Страница 44: ...ABP 90217A Operations Manual A 6 This page is intentionally blank...
Страница 45: ......