![Spacelabs Healthcare 90217A Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/spacelabs-healthcare/90217a/90217a_operation-manual_1339017011.webp)
1-5
070-2177-01 Rev B
1.3.4
Modem Speed and Compatibility Issues
The 90217A monitor operates Hayes-compatible modems only. The maximum
communication speed is 9600 baud. When the communications cable is plugged into the
monitor it should display 0999/9999
1.3.4.1 Connection Procedure
Connect the serial port cable (P/N 012-0096-01) between the 90217A monitor and the
modem.
Note:
If call waiting or call forwarding are options on a telephone used to transfer
data, ensure that both are deactivated. Otherwise modem communications
may be interrupted. In addition, telephone systems such as CBX or PBX can
cause interference with the modem, or the modem can cause interference
with the switching system..
To initialize the monitor for remote connection:
1.
Contact the base station by telephone (for remote operation only).
2.
Ask the base station operator to initialize the monitor. Give the following
information to the operator:
•
Patient’s name
• Patient ID number
• Whether monitor display is to be active or not
•
Time of day (12- or 24-hour format)
• Whether to display measurement (systolic/diastolic and heart rate)
•
Multiple or single cycle times. If using a single cycle for the 24-hour
monitoring period, indicate the cycle interval and whether the tone is ON
or OFF. For multiple cycle times, specify each cycle interval and whether the
tone is ON or OFF for each cycle.
•
Any other information the base station operator may request
The base station operator will enter the patient information in the computer.
3.
Prepare the monitor to receive the patient data from the base station.
• Power ON the modem.
• When instructed by the base station operator, power ON the ABP monitor.
Note:
The modem link must be established within 45 seconds. If this does not
occur, power the monitor OFF and then ON again to retry establishing the
link.
Содержание 90217A
Страница 4: ...ABP 90217A Operations Manual iv This page is intentionally blank...
Страница 24: ...ABP 90217A Operations Manual 1 18 This page is intentionally blank...
Страница 36: ...ABP 90217A Operations Manual 3 10 This page is intentionally blank...
Страница 44: ...ABP 90217A Operations Manual A 6 This page is intentionally blank...
Страница 45: ......