![Spacelabs Healthcare 90217A Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/spacelabs-healthcare/90217a/90217a_operation-manual_1339017020.webp)
ABP 90217A Operations Manual
1-14
“delicate” cycle only. Do not wash them in large commercial-type washers or with bed
linens or gowns.
Ensure that the carrying pouch is dry before re-use.
1.6.2
Removing/Installing the Bladder
To remove the bladder, follow the steps listed below:
1.
Using fingers only, fold or roll up the bladder inside the cuff. Do not use pencils,
pens, or other hard objects, because these may damage the bladder.
2. Remove the bladder through the hose exit opening.
Once the bladder is removed, mate together the hook-and-loop surfaces of the velcro
attachment before washing.
Re-install the bladder into the cuff wrap in reverse order. Make certain that all folds in the
bladder are removed and that the long end of the bladder fits into the long end of the cuff
(refer to Bladder configuration).
Note:
The cuff hose can exit from either opening in the cuff, depending on whether
you plan a right arm or left arm application.
1.7 Event Codes
The monitors display an event code whenever an event prevents the monitor from
successfully completing a blood pressure measurement. The two numerical digits of the
event code indicate the reason the measurement was aborted. The table below lists event
codes that are displayed on the monitor, as well as event codes that appear on the Event
Code Report.
Monitor Report
Condition
EC00
EC10
Excess movement artifact. Frequent EC10 messages may indicate an air
leak.
EC20
A) A very large number of movement artifacts
B) Heart rate arrhythmia
EC30
A) Movement artifact at mean arterial pressure
B) Heart rate arrhythmia
EC40
A) Movement artifact at systole
B) Heart rate arrhythmia
EC50
A) Movement artifact at diastole
B) Heart rate arrhythmia
Содержание 90217A
Страница 4: ...ABP 90217A Operations Manual iv This page is intentionally blank...
Страница 24: ...ABP 90217A Operations Manual 1 18 This page is intentionally blank...
Страница 36: ...ABP 90217A Operations Manual 3 10 This page is intentionally blank...
Страница 44: ...ABP 90217A Operations Manual A 6 This page is intentionally blank...
Страница 45: ......