![Spacelabs Healthcare 90217A Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/spacelabs-healthcare/90217a/90217a_operation-manual_1339017021.webp)
1-15
070-2177-01 Rev B
Monitor Report
Condition
EC60
A) Movement artifact
B) Heart rate arrhythmia
EC70
Systole was found to be above the highest cuff pressure. However, this
result appears to be an error caused by motion artifact. Therefore, the cuff
will not be inflated to a higher pressure on the next measurement attempt.
EC80
A) Movement artifact
B) Heart rate arrhythmia
EC90
A) Movement artifact
B) Heart rate arrhythmia
EC01
EC11
Did not pump above the mean arterial level.
EC21
Did not pump above systolic pressure.
EC91
Systole appears higher than the selected maximum cuff pressure limit.
EC02
EC12
Did not reach initial cuff pressure. The cuff may have been improperly
applied or there may be an air leak.
EC22
Overpressure.
EC32
Overpressure.
EC42
No cuff attached.
EC52
Kinked hose.
EC62
Cuff applied too loosely.
EC82
EC03
EC03
Patient canceled reading by pressing STOP key. No retry attempt is made
following an EC03 code.
EC13
The
Office Check
mode has been reinstated. No retry attempt is made
following an EC13 code.
EC04
EC04
Blood pressure measurement not completed in the maximum time allowed.
Occasional EC04 messages may result from excessive patient movement.
Frequent EC04 messages would indicate an improperly applied cuff or a
monitor malfunction which requires service.
ECn4
(where n = 1 to 9) Indicates that one or more of the blood pressure results
have been corrupted and subsequently recovered. Frequent occurrence of
this message would indicate a malfunction that requires service.
EC05
The individual blood pressure result has been corrupted and cannot be
recovered.
Содержание 90217A
Страница 4: ...ABP 90217A Operations Manual iv This page is intentionally blank...
Страница 24: ...ABP 90217A Operations Manual 1 18 This page is intentionally blank...
Страница 36: ...ABP 90217A Operations Manual 3 10 This page is intentionally blank...
Страница 44: ...ABP 90217A Operations Manual A 6 This page is intentionally blank...
Страница 45: ......